Medición de las propiedades eléctricas de la sangre humana utilizando un arreglo de microelectrodos coplanares
Abstract
This paper describes results in electrical properties measurement of human blood based on a platinum microelectrodes array, manufactured with electronic lithographic technique, and an impedance meter. The measurement frequency range was from 100 Hz up to 1 MHz and controlled blood temperature of 37 +- 0.5 C. According to with the experimental results in order to obtain the expected electrical properties valures, the polarization impedance effects must be taken into account. Such impedance effects were simulated in this work through an electrical model. The obtained data also show that for low measurement frequencies both the dielectric constant and the electric conductivity values of human blood are strongly influenced by the measurement cell geometry and parasite capacitances. The experimental results obtained by low and high measurement frequency values are compared with the results reported by other authors.
Downloads
Downloads
Published
How to Cite
Issue
Section
License
Upon acceptance of an article in the RMIB, corresponding authors will be asked to fulfill and sign the copyright and the journal publishing agreement, which will allow the RMIB authorization to publish this document in any media without limitations and without any cost. Authors may reuse parts of the paper in other documents and reproduce part or all of it for their personal use as long as a bibliographic reference is made to the RMIB. However written permission of the Publisher is required for resale or distribution outside the corresponding author institution and for all other derivative works, including compilations and translations.