We present a monitor of heart rate (HR) and electrocardiogram (ECG) represented in a TV tube. The device allows detecting ECG in the usual branches. It uses differential amplifiers with input impedance greater than 2.4 MOhms and common mode rejection greater than 80 dR. Its bandwidth is 0.05 to 100 Hz. The ECG is digitized and stored in memory for presentation in two beams with a 25 or 25 mm / s sweep. One of these strokes can be frozen for continuous analysis. The QRS is used to calculate the HR beat by beat, the result displayed on the screen along with the numbers represent the minimum and maximum HR to activate the audible alarm. Everything is controlled by a microprocessor.
Downloads
Downloads
Published
How to Cite
Issue
Section
License
Upon acceptance of an article in the RMIB, corresponding authors will be asked to fulfill and sign the copyright and the journal publishing agreement, which will allow the RMIB authorization to publish this document in any media without limitations and without any cost. Authors may reuse parts of the paper in other documents and reproduce part or all of it for their personal use as long as a bibliographic reference is made to the RMIB. However written permission of the Publisher is required for resale or distribution outside the corresponding author institution and for all other derivative works, including compilations and translations.